Clean seawater samples were collected from two sources; a GEOTRACES CTD referred to as GT-C/12L GoFlo, and the Super-GeoFISH towed surface vehicle. For more information on these methods, see the cruise report and the GeoFish Instrument Description.
Water column samples were filtered through pre-cleaned, 0.2 μm Pall Acropak Supor filter capsules as described elsewhere (e.g., Cutter & Bruland 2012; Hatta et al., 2015). Near surface water samples were collected using an underway towed-fish pumped seawater system (Bruland et al., 2005) with samples filtered through sequential 0.45 μm Osmonics and 0.2 μm Polycarbonate (PCTE) cartridge filters. Filtered water was collected in 125 mL HDPE bottles (Nalgene) that had been precleaned by soaking in hot 1.2 M HCl (reagent grade) for at least 8 h with subsequent thorough rinsing with ultrapure distilled deionized water (Barnstead E-pure). Samples were acidified in a laminar flow bench aboard ship using 0.5 mL of ultrapure HCl per 125 mL sample.
Dissolved Ga was determined by isotope dilution ICP-MS using a ThermoFisher Element 2 operated in low resolution. Samples were concentrated using Mg(OH)₂ co-precipitation (e.g., Shiller & Bairamadgi, 2006; Zurbrick et al., 2012). Briefly, in this technique, a small addition (~70 μL) of clean aqueous ammonia is added to the acidified seawater sample (~7.5 mL) which precipitates a fraction of the dissolved magnesium as the hydroxide, which in turn, scavenges the gallium from solution. An enriched isotope spike of known concentration was prepared using purified enriched ⁷¹Ga (99.8%), obtained from Oak Ridge National Laboratories.
Because there is a significant interference of doubly charged ¹³⁸Ba with ⁶⁹Ga, the precipitate was washed three times with a solution of high purity 0.1% NH₄OH to minimize residual Ba. The precipitate was then dissolved in 550 mL ultrapure 3% HNO3 (Seastar Chemicals, Baseline) and analyzed in low resolution using a ThermoFinnigan Element 2 High Resolution Inductively Coupled Plasma Mass Spectrometer (HR-ICP-MS). Isotopes monitored on the ICP-MS were ⁶⁹Ga, ⁷¹Ga, and ¹³⁸Ba. A slight correction for residual Ba was made based on the ratio of responses at masses 69 and 138 to a Ba standard solution. Because the residual salt content varied from sample to sample, it was not possible to matrix-match the Ba correction standard. However, typically, this correction affected the final result by < 2.5 pmol/kg; where higher Ba corrections were noted, the sample was reprecipitated and re-analyzed because of concerns about the accuracy of applying the Ba standard correction to samples of high salt content.
The reagent blank contribution to the dissolved Ga analysis is typically 0.6 pmol/kg and the detection limit (based on 3 times the standard deviation of the blank) is 0.3 pmol/kg. Repeated runs of US GEOTRACES intercalibration samples (GS and GD) and in-house reference solutions suggest a precision of +/- 4.0%; the limit of detection for Ga was 1.5 pmol/kg. Recovery of the method, as determined by repeated analysis of a spiked and unspiked seawater sample was 101.3 +/- 4.7%.