In order to ensure that 15N2 atom% effectuated by Membrane Inlet Mass Spectrometry, we tested whether the mass-specific signal ratios returned by MIMS are equivalent with parallel measurements on a Isotope Ratio Mass Spectrometer, on the basis that the latter returns accurate ratios (see 778000 cal_iso). MIMS and IRMS returned identical values for parallel samples.
Comparison of MIMS to IRMS
Granger, J., Bourbonnais, A., Wilson, S. (2019) Comparison of Membrane Inlet Mass Spectrometry (MIMS) to Isotope Ratio Mass Spectrometer (IRMS). Biological and Chemical Oceanography Data Management Office (BCO-DMO). (Version 1) Version Date 2019-10-02 [if applicable, indicate subset used]. doi:10.1575/1912/bco-dmo.778065.1 [access date]