The Tekran 2600 is a total Mercury (Hg) analysis system that can measure sub-picogram quantities of mercury in water, soil, vegetation, and other sample matrices. The system utilizes a Cold Vapour Atomic Fluorescence Spectroscopy (CVAFS) detector. The 2600-IVS (In-Vial Sparging) model is reconfigured for direct in-vial sparging sample introduction, and the 2600-NG (Natural Gas) model is designed for the analysis of gas phase sample cartridges. The system is capable of multiple method configurations: Dual stage gold preconcentration (EPA Method 1631); Direct measurement without preconcentration (EPA Method 245.7); Air sample analysis on gold traps (EPA Compendium Method IO-5); and Natural gas analysis on gold traps (ASTM D-6350, ISO 6978). It has a guaranteed minimum detection limit of < 0.05 nanograms per liter (ng/L). In clean room environments, with low mercury blanks, minimum detection limits as low as 0.02 ng/L are achievable.
Dataset Name | PI-Supplied Description | PI-Supplied Name |
---|---|---|
Dissolved mercury (Hg) speciation in the California Current System from samples collected on R/V Roger Revelle cruise RR2105 in July to August 2021 | Dissolved Total Mercury and elemental mercury were analyzed on a Tekran 2600. | Cold Vapor Atomic Fluorescence Spectroscopy (CVAFS) on a Tekran 2600 Automated Mercury Analyzer |
Suspended particle total mercury and monomethylmercury in the California Current Ecosystem (CCE) determined from samples collected on R/V Roger Revelle cruise RR2105 (P2107) in July to August 2021 | Cold vapor atomic fluorescence spectrometry (Tekran 2500/2600) |